Disclaimer: The information herein is not intended to replace the product’s User Manual. Please consult the User Manual for comprehensive information and cautionary guidance about the product’s operation and use.


Vscan Extend enters standby mode when the Power button is quickly pushed. In this mode, Vscan Extend will not have power to the probe or any display.

OR

Vscan Extend enters standby mode automatically while in freeze mode for more than 10 minutes.

Push the Power button again once to resume from standby mode. The device is ready to scan within 5 to 7 seconds.

NOTE: When resuming the device from standby mode, a new exam is created and the previous exam data is lost, if not previously saved.

 

Standby mode scenarios

  • When the battery capacity is > 90%, the system stays in standby mode for 3 hours and shuts down or when battery reaches 12% system shuts down.
  • When the initial battery capacity is between 90% and 50%, the system stays in standby mode for 2 hours and shuts down or when battery reaches 12% system shuts down.
  • When the initial battery capacity is between 50% and 25%, the system stays in standby mode for 1 hour and shuts down or when battery reaches 12% system shuts down.
  • When the initial battery capacity is between 25% and 12%, the system stays in standby mode for 15 minutes and shuts down or when battery capacity reaches 12% system shuts down.
  • When the initial battery capacity is < 12%, the system shuts down.