Cleaning The Device Print
Modified on: Tue, 11 May, 2021 at 2:28 AM
Disclaimer: The information herein is not intended to replace the product’s User Manual. Please consult the User Manual for comprehensive information and cautionary guidance about the product’s operation and use.
Make sure the device is disconnected from the AC/DC adapter before cleaning.
- Power off the Vscan Extend.
- Moisten a soft, non-abrasive cloth with a mild, general purpose, non-abrasive soap and water solution.
- Wipe the Vscan Extend.
- Wipe dry with a soft towel.
Reprocessing recommendation (Frequency)
After Each Use
- Inspect the transducer (see ‘Inspecting the Vscan Extend’)
- Clean the transducer (see ‘Disinfecting the probe’).
- If required, clean the device and the display (see ‘Using Germicides’).
- If required, disinfect the transducer (see ‘Disinfecting the probe’).
Ensure that the main unit and the transducer are properly cleaned after each use and before storage in the case.
Special Label Designations, Warnings and Precautions
Never use thinner, benzene, abrasive cleaners, or other strong solvents, as these may cause damage to the Vscan Extend.
Do not soak or saturate probes with solutions containing bleach, ammonium chloride compounds or hydrogen peroxide.
Special accessories
Special accessories are not applicable.
Point-of-Use Processing
No point-of-use processing is required prior to cleaning.
Cleaning Instructions
Cleaning the probe
- Remove the gel by wiping the probe lens with a soft cloth.
- Wipe the probe and cable with a soft cloth moistened with a warm soap and water solution (<80 0F/27 0C).
- Wipe the probe and cable with a soft cloth moistened with clean water (<80 0F/27 0C) until all soap is removed.
- Wipe dry with a soft towel.
Cleaning the USB door
- Open the USB door.
- Wipe the inside of the door with a soft cloth moistened with a mild, general purpose, non-abrasive soap and water solution.
- Wipe dry with a soft cloth.
Open the USB door
Did you find it helpful? Yes No
Send feedback